Restore Factory
Settings
All changes to the instrument defaults and settings can
be removed and the unit returned to its factory
settings.
From the Administrator menu touch
Restore
Factory Settings
.
Touch the
None
button and enter the administrator
password.
Touch
Restore Settings
and the unit will be reset
to factory defaults.
Touch
Cancel
to return to the Settings menu
without making any changes.
PROGRAM DEFAULTS
The program default settings will apply to every new program created on the Prime. However
individual programs can be edited as required.
Heated Lid
Use this to set the heated lid temperature to
approximately 10ºC higher than the hottest block
temperature in the program.
Touch
On/Off
to toggle between settings.
Touch the temperature button to set the required lid
temperature (35ºC to 115ºC).
Preheat lid
Set to
On
to pre-heat the lid to the set temperature
before the thermal cycling program begins.
Touch
On/Off
to toggle between settings.
Sample Cooling
Set to
On
to cool the block to 4ºC while the lid
preheats.
Touch
On/Off
to toggle between settings.
Touch
More Settings
for further program defaults
or
Back
to return to the Settings menu.
Pause
This will cause the unit to pause before the program
starts at the ambient temperature of the block unless
Sample Cooling has been set to On. In this case the
unit will pause with the block held at 4ºC and the
program will not commence until the unit is manually
un-paused.
Touch
On/Off
to toggle between settings.